Full Heat Disinfection System
When the MediQA Heat Disinfect Reverse Osmosis Machine is combined with the Heatsan Loop Heat Disinfection Unit, together they provide a full heat disinfect system that eliminates the need for routine, potentially dangerous chemical disinfection procedures.
Coupled with pre-treatment equipment, this fully integrated water purification system can accommodate facilities from 6 to 40 stations.
Reverse Osmosis Membrane Heat Disinfection
The MediQA Heat Disinfect Reverse Osmosis System useS proven reverse osmosis technology to provide a superior blend of reliability, efficiency and patient safety. With automated heat disinfection of RO membranes, MediQA units ensure that any potential for bacterial growth on membrane surfaces is minimized.
User-friendly operation is managed through a full color touch screen display with comprehensive menu navigation – including full schematics. Password protected stored data includes active status indicators, alarm conditions and performance logs. The stored data can be downloaded to a remote PC or accessed through the touch screen.
Systems with single or dual pass membranes are available.
Distribution Loop Heat Disinfection
The Heatsan is an integrated heat disinfection system for the distribution loop with a significant advantage: the ability to adjust operation for each day of the week, heating intervals, holding times and disinfection temperatures. As a result, the Heatsan can save energy and staff time while providing up to 100 gallons of 185°F water for disinfection.
Incorporating a PLC controller, the Heatsan is operated by a dual color touch screen displaying alarms, system status, tank level and disinfection temperatures plus a password-protected event log holding up to 12 months of data. The construction is rugged yet space-efficient, featuring a stainless steel skid and an insulated stainless steel storage tank with a sloping base for easy drainage.
Reverse Osmosis Features:
- Single and dual pass membrane configurations
- Provides maximum flexibility in water quality
- Automated heat disinfection of RO membranes
- Provides system efficiency and labor savings
- Ensures that any potential for bacterial growth on membrane surfaces is minimized
- Full color touch screen display with comprehensive menu navigation, including full schematics
- Advanced data logging technology with password protection
- Information available via an integrated touch screen display or downloaded to Excel
- Stored data includes active status indicators, alarm conditions and performance logs
- 8” TFC membranes in FRP vessels for durability
- Compact, powder coated aluminum skid mounted package
Loop Heat Disinfect Features:
- Easy step through programming (7 day program)
- Program syncs dialysis machines with the Heatsan
- Intelligent monitoring routines, leak detection, and over use
- Heat disinfect will abort if leak is detected
- Four levels of password protection
- Event log holds up to 12 months of data including AAMI standards required end of loop history
- USB download to Excel
- Stainless steel skid complete with insulated stainless steel storage tank with sloping base fitted with three heating elements to minimize electrical consumption, level transmitter and temperature monitor
- Stainless steel recirculation pump, stainless steel solenoid valves
- The isolation of the RO system is accomplished with three way motorized ball valves that are controlled by the Heatsan
MediQA RO Machine Specifications: | ||||
Membrane Type: | ||||
Heat Disinfectable TFC | Recovery Rate: 50-75% | |||
Salts Rejection: >96% | pH Tolerance: 2-11 | |||
Feed Water Requirements: | ||||
Total Dissolved Solids: <1000 ppm | Total Chlorine: < 0.1 ppm | |||
Total Hardness: < 4 ppm at CaCO3 | Fouling Index (SDI): < 5 | |||
Feed Pressure (PSI): 20 – 90 | Temperature: 41 – 95°F | |||
Connections: | ||||
Feed Water: 1.25” FNPT | Overflow Tank Drain: 1.5″ FNPT | |||
Concentrate (reject): 1″ FNPT | Permeate (product): 1″ FNPT | |||
Loop Return: 1″ FNPT | ||||
Single Pass | Dual Pass | |||
Model Number: | MSP2 | MSP3 | MDP2 | MDP4 |
Gallons/Day* (up to): | 12,960 | 17,280 | 7,920 | 14,440 |
Gallons/Minute* (up to): | 9 | 12 | 5.5 | 10 |
Number of Membranes: | 2 | 3 | 3 | 5 |
Replacement Membrane: R22-8041 (8″ X 40″) | ||||
Minimum Feed Flow: | 26GPM | 29GPM | 29GPM | 30GPM |
Power Requirements: | ||||
Voltage: 208/230V, 60 Hz, 3 phase with dedicated neutral and ground | ||||
Max Current Draw (AMP) | 44 | 44 | 44 | 44 |
Shipping Weight Lbs: | 1,500 | 1,500 | 1,820 | 2,470 |
Operation Weight Lbs: | 1,310 | 1,375 | 1,660 | 2,285 |
Dimensions (W x D x H): | 43 x 64 x 79 | 43 x 76 x 79 | ||
* Permeate flow rate at 77°F. Variations in feed water temperature can affect output by up to ±3% per degree Fahrenheit. Based on 1,000 microsiemens water or less. This sheet is not to be used for sizing purposes. Contact an AmeriWater representative and provide a loop design to properly size the system. | ||||
Heatsan Specification: | ||||
Models | HS208 | HS460 | HS575 | |
Electrical | 208v, 32 Amps 3 Ph | 460V, 16 Amps, 3 Ph | 575V, 13 Amps, 3 Ph | |
Performance | Tank Volume | 100 Gallons (380 Liters) | ||
Pump Output Range | 5.2 GPM @ 80 PSI 19.3 GPM @ 30 PSI | |||
Designed Max Water Temperature | 203°F | |||
Max. No. of Dialysis Machines | Depends on the volume of hot water required per machine and the rate at which they draw water. Total volume required not to exceed 95 gallons. Draw rate not to exceed pump performance. | |||
Connections In | Drain | 1½ Sanitary Tri-clamp Coupling | ||
Overflow | ||||
Loop IN/OUT | ||||
Fill | ½ Sanitary Tri-clamp Coupling | |||
Hot Water Flow | 1½ Sanitary Tri-clamp Coupling | |||
Hot Water Return | ||||
Dimensions | 28 x 35 x 70 | |||
Weight Lbs. (Shipping/Operating) | 700 / 1,580 |
- Entire device conforms to IEC UL standard 61010-1
- FDA 510K / ISO 13485 Registered Medical Device
- Health Canada Medical Device License 90626
- Health Canada Medical Device License 90655